Essential Dentist Skills: Manual Cleaning Tips and Procedures for Dental Handpieces

Dental handpiece is an essential tool in stomatology treatment. A large number of studies have shown that since most oral clinical operations are performed in the patient’s mouth, the dental handpiece frequently contacts the patient’s saliva, blood, and oral tissues during operation, coupled with the fine structure of the handpiece, it is tightly sealed by a metal casing, and the interior is complicated The turbine bearings, water and gas pipelines, and cavities of handpieces are often a medium for cross infection.
There are three main ways that dental handpieces can cause cross infection.
1. During the operation of the handpiece in the oral cavity, the surface contamination caused by the patient’s saliva, blood, and dandruff, and the internal pollution of the handpiece penetrating into the interior of the handpiece through the gap of the handpiece itself.
2. Air pollution caused by aerosols and droplets with pathogenic microorganisms generated during high-speed rotary cutting of handpieces.
3. The negative pressure formed when the handpiece’s high-speed turbine stops rotating can suck back the pathogenic microorganisms in the patient’s mouth into the handpiece and enter the integrated treatment station water and gas pipeline system through the handpiece connector, resulting in pollution, or due to accumulation. Bacteria or harmful microorganisms in the water and gas pipelines of the comprehensive treatment platform are breeding and pollute the pipelines.
Dental handpiece manual cleaning process and key points
1. Rinse: Rinse the handpiece under running water to remove the surface pollutant.
2 . Brushing: Soak for 2-5min with multiple enzymes, use a soft brush to brush under the liquid surface, focusing on scrubbing the head of the handpiece.
3. Rinse, terminal rinse: Rinse or brush under running water. The lumen was flushed with a pressure water gun. Final rinse with pure or soft water.
4. Disinfection: physical disinfection, water temperature above 90 ° C, time ≥1min. Change gloves after disinfection. Do not soak the handpiece in the disinfectant solution, so as not to corrode the parts inside the handpiece.
Change gloves after disinfection. Do not soak the handpiece in the disinfectant solution, so as not to corrode the parts inside the handpiece.
5. Drying: Use a pressure air gun or a syringe to blow dry the moisture on the surface of the pipe and the bearing of the rotor, or inject 75% ethanol to dry.
6. Oil injection: mechanical oil injection: use full automatic oil injection equipment to inject special handpiece cleaning lubricant into the handpiece.
Manual oil injection: Spray type cleaning lubricant should be used.
7. Inspection: Check the cleanliness and performance of handpieces and equipment after oiling.
8. Drying: drying cabinet drying, temperature 70-90 ℃, time 5-10min
9. Inspection: check the cleanliness and performance of the handpiece after visual inspection or a magnifying glass with a light source