Planting Handpiece Maintenance Precautions

After each use, the handpiece must be cleaned and oiled. It is recommended to use a cleaning and oiling machine for cleaning and oiling. This effect is the best and the service life will be longer. If there is no oil filling machine, you must use the manual oiling method. When using manual oiling, be sure to install an oiling pipe at the end of the handpiece to fill the oil. The time for manual oiling is about 5 seconds. Be sure to see the front of the machine head overflow with clean oil.

  1. After cleaning and filling, be sure to put it in a sterilization bag for high temperature sterilization. Never put it under high temperature and high pressure without putting it in a sterilization bag.
  2. Never soak in glutaraldehyde or a liquid containing disinfectant.
  3. The motor of the planter must not enter water or oil. If the water enters, the motor must be dried.
  4. If the motor is to be autoclaved, it is necessary to install a motor closing cap on the mouth of the motor to prevent water from entering the motor. At the same time, the motor must be sterilized in a sterilization bag to prevent water from entering the handpiece.
  5. Planting handpieces must be cleaned and disinfected with oil after each use. If the handpiece is not used for a long time, it must be filled with oil and cleaned before use.